98
Keysight E1412A Digital Multimeter Service Guide
E1412A Multimeter Specifications
Frequency and Period Characteristics
Frequency and Period Characteristics
Accuracy Specifications
(% of reading) [1] [4]
Additional Low-Frequency Errors
(% of reading) [4]
Integration Time (number PLCs)
Frequency
100 & 10
1 & 0.2
0.02
3Hz-5Hz
0
0.12
0.12
5Hz-10Hz
0
0.17
0.17
10Hz-40Hz
0
0.2
0.2
40Hz-100Hz
0
0.06
0.21
100Hz-300Hz
0
0.03
0.21
300Hz-1kHz
0
0.01
0.07
>1kHz
0
0
0.02
Measuring Characteristics
Frequency and Period
Measurement Method:
Reciprocal-counting technique. AC-coupled input using the ac
voltage
measurement function.
Voltage Ranges:
100 mV rms full scale to 300V rms. Auto or manual ranges.
Gate Time:
10 ms, 100 ms or 1 second.
Settling Considerations
Errors will occur when attempting to measure the frequency or period of an input following a dc
offset voltage change. The input blocking RC time constant must be allowed to adequately settle (up
to 1 second) before the most accurate measurements are possible.
Measurement Considerations
All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals.
Shielding inputs from external noise pickup is critical for minimizing measurement errors.
Function
Range [3]
Frequency
24 Hour [2]
23
C
1
C
90 Day
23
C
5
C
1 Year
23
C
5
C
Temperature
Coefficient
0
C - 18
C
28
C - 55
C
Frequency,
Period
100mV
to
300V
3Hz - 5Hz
5Hz - 10Hz
10Hz - 40Hz
40Hz - 300kHz
0.10
0.05
0.03
0.006
0.10
0.05
0.03
0.01
0.10
0.05
0.03
0.01
0.005
0.005
0.001
0.001
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......