E1412A 6 1/2 Digit Multimeter Service Guide
Keysight E1412A Multimeter Error Messages
Execution Errors
-109 Missing parameter
Fewer parameters were received than expected for
the command. You omitted one or more parameters that are required for this
command. Example:
SAMP:COUN
-112 Program mnemonic too long
A command header was received which
contained more than the maximum 12 characters allowed. Example:
CONFIGURATION:VOLT:DC
A <code> string contains more than the maximum 12 characters allowed in the
CALibration:SECure:CODE command.
-113 Undefined header
A command was received that is not valid for this
multimeter. You may have misspelled the command or it may not be a valid
command. If you are using the short form of the command, remember that it may
contain up to four letters. Example: TRIGG:COUN 3
-121 Invalid character in number
An invalid character was found in the
number specified for a parameter value. Example: STAT:QUES:ENAB
#B01010102
-123 Numeric overflow
A numeric parameter was found whose exponent was
larger than 32,000. Example: TRIG:COUN 1E34000
-124 Too many digits
A numeric parameter was found whose mantissa
contained more than 255 digits, excluding leading zeros.
-128 Numeric data not allowed
A numeric parameter was found but a
character string was expected. Check the list of parameters to verify you have
used a correct parameter type. Example: TRIG:SOUR 1
-131 Invalid suffix
A suffix was incorrectly specified for a numeric parameter.
You may have misspelled the suffix. Example: TRIG:DEL 0.5 SECS
-138 Suffix not allowed
A suffix was received following a numeric parameter
which does not accept a suffix. Example: SAMP:COUN 1 SEC (SEC is not a valid
suffix).
-148 Character data not allowed
A character string was received but a
numeric parameter was expected. Check the list of parameters to verify that you
have used a valid parameter type. Example: CAL:LFR XYZ
-151 Invalid string data
An invalid character string was received. Check to
see if you have enclosed the character string in single or double quotes.
Example: CAL:STR ’NEXT CAL DUE 10/4/1996(the ending quote is missing).
-158 String data not allowed
A character string was received but is not
allowed for the command. Check the list of parameters to verify that you have
used a valid parameter type. Example: CALC:STAT ’ON’
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......