Keysight E1412A 6 1/2 Digit Multimeter Service Guide
23
Quick Verification Tests
Verification Tests
Quick Verification Tests
The Quick Verification procedure is a combination of internal self-test and an
abbreviated set of the Performance Verification Tests that follow (specified by the
letter Q in the performance verification test tables). This test provides a simple
method to achieve high confidence in the meter’s ability to functionally operate
and meet specifications. These tests represent the absolute minimum set of
performance tests that are recommended following any service activity. These
tests are designed to provide a moderate level of confidence that the instrument
is meeting specifications.
Perform the Quick Verification tests by doing the following:
–
Perform a compete self-test using the *TST? command.
–
Perform the performance verification tests indicated with the letter Q.
If your multimeter fails Quick Verification, recalibrate the section that fails. For
example, if ac voltage fails, it may be because the internal shields have been
removed and reinstalled. Recalibrate the ac voltage function only and reverify the
performance of ac voltage.
Carefully review the test conditions at the beginning of the chapter for possible
causes of performance verification test failures. Take into consideration all test
conditions, your test set-up and the handling of test connectors and shorting
devices. Contact your nearest Keysight Technologies service office if your
multimeter fails to meet specification.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......