50
Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Adjustments
Adjustment Procedures
7
Repeat steps 3 through 6 for the ranges and inputs given in the following
table.
DC Voltage Function Adjustment Points
Keysight E1412A Range
DC Standard Output
1
1 Valid calibration input values sent to the multimeter are 0.9 to 1.1 of Full
Scale for the range being adjusted (270V to 300V for the 300V range).
The <input voltage> parameter of the CAL:VAL command must equal the
actual input value. For example, if you input 0.9V to calibrate the 1V
range, you send CAL:VAL 0.9 to the multimeter prior to CAL?.
100 mV
100 mV
1 V
1 V
10 V
10 V
100 V
100 V
300 V
300 V
Figure 5-3
DC Voltage Gain Adjustment Setup.