94
Keysight E1412A Digital Multimeter Service Guide
E1412A Multimeter Specifications
AC Characteristics
AC Characteristics
Accuracy Specifications
(% of r % of range)
[1]
Function
Range
[3]
Frequency
24 Hour
[2]
23
C
1
C
90 Day
23
C
5
C
1 Year
23
C
5
C
Temperature
Coefficient
0
C - 18
C
28
C - 55
C
True RMS
AC
Voltage
[4]
100.0000 mV 3 Hz-5 Hz
5 Hz-10 Hz
10 Hz-20 kHz
20 kHz-50 kHz
50 kHz-100 kHz
100 kHz-300 kHz
1.00 + 0.03
0.35 + 0.03
0.04 + 0.03
0.10 + 0.05
0.55 + 0.08
5.00 + 0.50
1.00 + 0.04
0.35 + 0.04
0.05 + 0.04
0.11 + 0.05
0.60 + 0.08
5.00 + 0.50
1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.05
0.60 + 0.08
5.00 + 0.50
0.100 + 0.004
0.035 + 0.004
0.005 + 0.004
0.011 + 0.005
0.060 + 0.008
0.200 + 0.020
1.000000 V
to
100.000V
[12]
3 Hz-5 Hz
5 Hz-10 Hz
10 Hz-20 kHz
20 kHz-50 kHz
50 kHz-100 kHz
100 kHz-300 kHz
1.00 + 0.02
0.35 + 0.02
0.04 + 0.02
0.10 + 0.04
0.55 + 0.08
5.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.60 + 0.08
5.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.60 + 0.08
5.00 + 0.50
0.100 + 0.003
0.035 + 0.003
0.005 + 0.003
0.011 + 0.005
0.060 + 0.008
0.200 + 0.020
300.000V
[12]
3 Hz-5 Hz
5 Hz-10 Hz
10 Hz-20 kHz
20 kHz-50 kHz
50 kHz-100 kHz [5]
100 kHz-300 kHz
[5]
1.00 + 0.06
0.35 + 0.06
0.04 + 0.06
0.10 + 0.12
0.55 + 0.24
5.00 + 1.50
1.00 + 0.09
0.35 + 0.09
0.05 + 0.09
0.11 + 0.15
0.60 + 0.24
5.00 + 1.50
1.00 + 0.09
0.35 + 0.09
0.06 + 0.09
0.12 + 0.15
0.60 + 0.24
5.00 + 1.50
0.100 + 0.009
0.035 + 0.009
0.005 + 0.009
0.011 + 0.015
0.060 + 0.024
0.200 + 0.060
True RMS
AC
Current
[4]
1.000000 A
3 Hz-5 Hz
5 Hz-10 Hz
10 Hz-1 kHz
1 kHz-5 kHz
1.05 + 0.04
0.35 + 0.04
0.15 + 0.04
0.40 + 0.04
1.05 + 0.04
0.35 + 0.04
0.15 + 0.04
0.40 + 0.04
1.05 + 0.04
0.35 + 0.04
0.15 + 0.04
0.40 + 0.04
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.015 + 0.006
3.00000 A
3 Hz-5 Hz
5 Hz-10 Hz
10 Hz-1 kHz
1 kHz-5kHz
1.70 + 0.06
0.95 + 0.06
0.75 + 0.06
1.00 + 0.06
1.70 + 0.06
0.95 + 0.06
0.75 + 0.06
1.00 + 0.06
1.70 + 0.06
0.95 + 0.06
0.75 + 0.06
1.00 + 0.06
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.015 + 0.006
Additional Low Frequency Errors
(% of reading)
Frequency
AC Filter
3 Hz
20 Hz
200 Hz
10Hz-20Hz
0
0.74
--
20Hz-40Hz
0
0.22
--
40Hz-100Hz
0
0.06
0.73
100Hz-200Hz
0
0.01
0.22
200Hz-1kHz
0
0
0.18
>1kHz
0
0
0
Additional Crest Factor Errors
(non-sinewave)
[6]
Crest Factor
Error (% or reading)
1-2
0.05%
2-3
0.15%
3-4
0.30%
4-5
0.40%
NOTE: Crest Factor is not specified for
non sinewave inputs <100Hz using the
slow (3Hz) AC filter. See note [6].
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......