Keysight E1412A Digital Multimeter Service Guide
95
AC Characteristics
E1412A Multimeter Specifications
Measuring Characteristics
Measurement Noise Rejection
[7]
AC CMRR
70 dB
True RMS AC Voltage
Measurement Method:
AC-coupled True RMS - measures the ac component of the input with
up to
300Vdc of bias on any range (max AC + DC = 300Vrms).
Crest Factor:
Maximum 5:1 at full scale
AC Filter Bandwidths:
3Hz-300kHz (Slow filter)
20Hz-300kHz (Medium filter)
200Hz-300kHz (Fast filter)
Input Impedance:
1M
2%, in parallel with 100pF
Input Protection:
300Vrms all ranges
True RMS AC Current
Measurement Method:
Direct coupled to the fuse and shunt. AC-coupled True RMS
measurement
(measures the ac component only).
Shunt Resistor:
0.1
for 1A and 3A ranges
Burden Voltage:
1A range: <1Vrms
3A range: < 2Vrms
Input Protection:
Externally accessible 3.15A, 250V, Class H fuse
Class H fuses are fuses with a high interrupt rating which defines a
fuse's ability to safely interrupt and clear short circuits. Replace the
fuse with part number 2110-0957, (3.15A, 250V, 5.0 mm diameter,
20.0 mm long) or use Cooper Industries Inc. fuse part number
GDA-3.15.