Keysight E1412A Digital Multimeter Service Guide
101
To Calculate Total Measurement Error
E1412A Multimeter Specifications
To Calculate Total Measurement Error
Each specification includes correction factors which account for errors present
due to operational limitations of the multimeter. This section explains these
errors and shows how to apply them to your measurements. Refer to the section
titled “Interpreting Multimeter Specifications” to get a better understanding of
the terminology used and to help you interpret the multimeter's specifications.
The multimeter's accuracy specifications are expressed in the form: (
% of
reading
+
% of range
). In addition to the reading error and range error, you may
need to add additional errors for certain operating conditions. Check the list
below to make sure you include
all
measurement errors for a given function. Also,
make sure you apply the conditions as described in the footnotes on the
specification pages.
-
If you are operating the multimeter outside the 23°C
5°C temperature range
specified, apply an additional
temperature coefficient error
.
-
For dc voltage, dc current, and resistance measurements, you may need to
apply an additional
reading speed error
or
autozero OFF error
.
-
For ac voltage and ac current measurements, you may need to apply an
additional
low frequency error
or
crest factor error
.
Understanding the “% of reading” Error
The
reading error
compensates for inaccuracies that result from the function and
range you select, as well as the input signal level. The reading error varies
according to the input level on the selected range. This error is expressed in
percent of reading. The following table shows the reading error applied to the
multimeter's 24-hour dc voltage specification.
Range
Input Level
Range Error
(% of range)
Range Error
Voltage
10Vdc
10Vdc
0.0015
150
V
10Vdc
1Vdc
0.0015
15
V
10Vdc
0.1Vdc
0.0015
1.5
V
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......