Keysight E1412A Digital Multimeter
Service Guide
91
A
E1412A Multimeter Specifications
DC Characteristics
Accuracy Specifications
(% of r % of range) [1]
Function
Range
[3]
Test
Current
or
Burden
Voltage
24 Hour
[2]
23
C
1
C
90 Day
23
C
5
C
1 Year
23
C
5
C
Temperature
Coefficient
0
C - 18
C
28
C - 55
C
DC Voltage
100.0000mV
1.000000V
10.00000V
100.0000V
300.0000V
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0018
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0030
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0030
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0003
Resistance
[4]
100.0000
1.000000k
10.00000k
100.0000k
1.000000M
10.00000M
100.0000M
1mA
1mA
100
A
10
A
5
A
500nA
500nA ||
10M
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0010
0.0150 + 0.0010
0.3000 + 0.0100
0.0080 + 0.0040
0.0080 + 0.0010
0.0080 + 0.0010
0.0080 + 0.0010
0.0080 + 0.0010
0.0350 + 0.0010
0.8000 + 0.0100
0.0100 + 0.0040
0.0100 + 0.0010
0.0100 + 0.0010
0.0100 + 0.0010
0.0100 + 0.0010
0.0540 + 0.0010
0.8000 + 0.0100
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
DC Current
10.00000mA
100.0000mA
1.000000A
3.000000A
<0.1V
<0.7V
<1V
<2V
0.0050 + 0.0100
0.0100 + 0.0040
0.1000 + 0.0060
0.7000 + 0.0200
0.0500 + 0.0200
0.0500 + 0.0050
0.1300 + 0.0100
0.7200 + 0.0200
0.0700 + 0.0200
0.0700 + 0.0050
0.1500 + 0.0100
0.7200 + 0.0200
0.0050 + 0.0020
0.0060 + 0.0005
0.0060 + 0.0010
0.0060 + 0.0020
DC:DC Ratio
100mV to
300V
(Input Accuracy) + (Reference Accuracy)
Input Accuracy = accuracy specification for the HI-LO input signal.
Reference Accuracy = accuracy specification for HI-LO reference
input signal (Sense HI-LO input terminals).
NOTE:
Autorange is used for the reference signal regardless of the range set for the
HI-LO input signal. The 10V range is the highest range available for the
reference signal and the highest range the multimeter will autorange to for
measuring the reference signal.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......