Keysight E1412A Digital Multoimeter User Guide
ccxxv
Execution Errors
Keysight E1412A Multimeter Error Messages
-148
Character data not allowed
A character string was received but a numeric parameter was expected. Check
the list of parameters to verify that you have used a valid parameter type.
Example:
CAL:LFR XYZ
-151
Invalid string data
An invalid character string was received. Check to see if you have enclosed the
character string in single or double quotes. Example:
CAL:STR 'NEXT CAL DUE
10/4/1996
(the ending quote is missing).
-158
String data not allowed
A character string was received but is not allowed for the command. Check the
list of parameters to verify that you have used a valid parameter type.
Example:
CALC:STAT 'ON'
-160
to
-168
Block data errors
The multimeter does not accept block data.
-170
to
-178
Expression errors
The multimeter does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (GET) or
*TRG
was received but the trigger was ignored.
Make sure the multimeter is in the “wait-for-trigger” state before issuing a
trigger, and make sure the correct trigger source is selected.
-213
Init ignored
An
INITiate
command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a
measurement in progress and place the multimeter in the “idle” state.
-214
Trigger deadlock
A trigger deadlock occurs when the trigger source is
BUS
and a
READ?
command is received.
-221
Settings conflict
This error can be generated in one of the following situations:
You sent a
CONFigure
or
MEASure
command with autorange enabled and with a
fixed resolution. Example:
CONF:VOLT:DC DEF,0.1