Keysight E1460A 64-Channel Relay Multiplexer User Guide
73
ABORt
Relay Multiplexer Command Reference
ABORt
The ABORt command stops a scan in progress when the scan is enabled via the
interface and the trigger source is TRIGger:SOURce BUS or TRIGger:SOURce
HOLD.
Subsystem Syntax
ABORt
Comments
ABORt Actions:
The ABORt command terminates a scan in progress by causing the
switchbox to no longer wait for a trigger. When the ABORt command is
executed, the last channel switched during the scan remains in the position.
Stopping Scan Enabled Via Interface:
When a scan is enabled via an interface, an
interface clear command (CLEAR 7) can be used to stop the scan. When the scan
is enabled via the interface and TRIG:SOUR BUS or HOLD is set, you can use
ABORt to stop the scan.
Restarting a Scan:
Use the INITiate command to restart the scan.
Related Commands:
ARM, INITiate:CONTinuous, [ROUTe:]SCAN, TRIGger
Example
Stopping a Scan with ABORt
This example stops a (continuous) two-wire scan in progress in a single-module
switchbox.
TRIG:SOUR BUS !Trigger command will be via backplane (bus) interface
(*TRG command generates trigger)
INIT:CONT ON !Set continuous scanning
SCAN (@100:107) !Scan channels 0 to 7 in bank 0
INIT !Start scan, close channel 0
.
.
.
ABOR !Abort scan in progress