Data logger properties and waveform settings
Items
Available settings
Description
1
Vertical Settings
(Vr)
Trace checkbox
Selects the trace to display for the output. If no checkbox is enabled, data logging will not
logged for the selected output.
Configures the voltage/division, current/division, power/division and offsets for each out-
put.
2
Horizontal Set-
tings (Hr)
Time/Division
Configures to zoom in or out of the data so that you can view waveform details. The
numbers on the top of the display indicate the location of the data that is being viewed
relative to the entire data log. Specified in time/division on the x axis. Applies to ALL
traces.
Offset
Configures the waveform location (right or left) of the horizontal offset reference. The trig-
ger point is indicated by the solid arrow
.
3
Logging (Lg)
Duration
Configures the duration of the data log in hours, minutes, and seconds. The maximum
duration is 30,000 hours. The logging information applies to data logger measurements on
all outputs.
Sample Period
Configures the interval between data samples in milliseconds, which can be set from 200
milliseconds to 60 seconds.
Trigger Source
Configures the trigger source for data logger.
Trigger Pos
Trigger Position specifies a trigger offset. This allows the specified time of pre-trigger
data to be logged to the file. For example, if you specified a data log duration of 30 s and a
trigger position of 50%, the Data Logger will log 15 s of pre-trigger data to the file before
the trigger occurs. Subsequently, 15 s of post-trigger data will then be logged to the data
file.
Log Min/Max checkbox When enabled, lets you log the minimum and maximum values of each sample to the data
log file. With Log Min/Max checked, the resulting file size will triple.
Menu
Available settings
Description
File Selection
-
Configures the file location and filename for logged data, export the file to .csv format and
load the previously logged file to the instrument.
Preset
–
Returns the Data Logger View to the power-on display settings.
Settings
Vr, Hr or Lg
Selects the Vertical Settings, Horizontal Settings or Logging Settings field.
Action required:
Press Settings to toggle between Vr, Hr and Lg.
100