Settings
Available Key Settings
Description
Voltage /Current
After List
DC or List
Specifies what happens when the output sequence completes.
Default is DC.
Action required:
Press V/I List to toggle between DC and List.
DC (Return to DC
Value)
Returns to the DC value that was in effect before the output sequence
started.
List (Last List Value) Remains at the last list value.
Pace
Dwl or Trg
Configures the pacing of the step.
Default is Dwl.
Action required:
Press Pace to toggle between Dwl and Trg.
Dwl
(Dwell)
The next step immediately outputs when the dwell time has elapsed.
Trg (Trigger)
The next step immediately outputs when an external trigger is received.
If the step time completes before trigger occurs, the step remains at the
last list value while waiting for the trigger.
Voltage Mode
Fix, Stp or Lst
Sets the voltage mode. This determines what happens to the output voltage when the system is
initiated or triggered.
Default is Fix
Note: If current mode is set to Step, voltage mode is not allowed to set to List and vice versa.
Action required:
Press V Mode to toggle between Fix, Stp and Lst.
Fix (Fixed)
Keeps the output at its immediate value.
Stp (Step)
Steps the output to the triggered level when a trigger occurs
Lst (List)
Causes the output to follow the list values when a trigger occurs.
Current Mode
Fix, Stp or Lst
Sets the current mode. This determines what happens to the output current when the system is
initiated or triggered.
Default is Fix
Note: If voltage mode is set to Step, current mode is not allowed to set to List and vice versa.
Action required:
Press I Mode to toggle between Fix, Stp and Lst.
Fix (Fixed)
Keeps the output at its immediate value.
Stp (Step)
Steps the output to the triggered level when a trigger occurs
Lst (List)
Causes the output to follow the list values when a trigger occurs.
Keysight E36300 Series User's Guide
77