Test considerations
– Ensure that the calibration ambient temperature is stable and between 20 °C and 30 °C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Keep cables as short as possible, consistent with the impedance requirements.
– For E36311A, performance verification and calibration procedure must be performed through front panel output.
– For E36312A and E36313A, performance verification and calibration procedure must be performed through rear
panel output.
The tests should be performed by qualified personnel. During performance verification tests, hazardous voltages
may be present at the outputs of the power supply.
Measurement techniques
Voltmeter
To ensure that the values read by the voltmeter during both the verification procedure and the calibration procedure
are not affected by the instantaneous measurement of the AC peaks of the output current ripple, make several DC
measurements and average them.
Current-monitoring resistor
The 4-terminal current shunt is used to eliminate output current measurement error caused by voltage drops in the
load leads and connections. It has special current-monitoring terminals inside the load connection terminals.
Connect the voltmeter directly to these current-monitoring terminals.
Electronic load
Many of the test procedures require the use of a variable load capable of dissipating the required power. If a variable
resistor is used, switches should be used to connect, disconnect, or short the load resistor. For most tests, an
electronic load can be used. The electronic load is considerably easier to use than load resistors, but it may not be
fast enough to test transient recovery time and may be too noisy for the noise (PARD) tests.
Fixed load resistors may be used in place of a variable load, with minor changes to the test procedures. Also, if
computer controlled test setups are used, the relatively slow (compared to computers and system voltmeters)
settling times and slew rates of the power system may have to be taken into account. "Wait" statements can be used
in the test program if the test system is faster than the power system.
Keysight E36300 Series Service Guide
23