Calibration procedure
Enter the calibration state
To begin the calibration procedure, you must enter the calibration state.
Step
Front Panel
SCPI
1
Press Utilities > Test / Setup > Calibration.
Enter the default passcode (default passcode is 0). Press Login to enter the cal-
ibration page.
CAL:SEC:STAT 0, <
code
>
2
To change a forgotten passcode to a new passcode, you can turn on the instrument
after shorting CAL SECURE jumper J10 on the front panel board and send scpi com-
mand to change the passcode.
CAL:SEC:CODE <code>
Calibrate voltage
Let the unit sit with output ON for one minute, then connect the DMM voltage input to the power supply.
Step
Front Panel
SCPI
1
Press Perform Calibration and select the output for calibration. Press Cal
Volt.
CAL:VOLT:LEV MIN, (@<channel>)
2
Measure the output voltage (low point) with the DMM.
3
Enter the measured value, and press Next.
CAL:VOLT <
reading
>, (@<channel>)
CAL:VOLT:LEV MAX, (@<channel>)
4
Measure the output voltage (high point) with the DMM.
5
Enter the measured value, and press Next.
CAL:VOLT <
reading
>, (@<channel>)
6
Read DONE or FAIL on the display.
(wait 30 seconds)
SYST:ERR?
7
Press Cal Save.
CAL:SAVE
Keysight E36300 Series Service Guide
45