Calibrate current
Let the unit sit with output ON for one minute, then connect a current monitoring resistor across the output
terminals to be calibrated and then connect a DMM across the terminals of the monitoring resistor.
Step
Front Panel
SCPI
1
Press Perform Calibration and select the output for calibration. Press Cal
Curr.
CAL:CURR:LEV MIN, (@<channel>)
2
Measure the output current (low point) with the DMM.
3
Enter the measured value, and press Next.
CAL:CURR <
reading
>, (@<channel>)
CAL:CURR:LEV MAX, (@<channel>)
4
Measure the output current (high point) with the DMM.
5
Enter the measured value, and press Next.
CAL:CURR <
reading
>, (@<channel>)
6
Read DONE or FAIL on the display.
(wait 30 seconds)
SYST:ERR?
7
Press Cal Save.
CAL:SAVE
Calibrate low range current (For E36312A and E36313A only)
Allow the unit to sit with output ON for one minute before continuing.
Step
Front Panel
SCPI
1
Press Perform Calibration and select the output for calibration. Press Cal
Low Curr.
CAL:CURR:LEV:LOW MIN, (@<channel>)
2
Wait approximately 50 seconds. Connect calibration resistor with the correct value as stated in the instruction shown on the display.
Measure the output low range current (low point) with the DMM.
3
Enter the measured value, and press Next.
CAL:CURR:LOW <reading>, (@<channel>)
4
Measure the output low range current (high point) with the DMM.
CAL:CURR:LEV:LOW MAX, (@<channel>)
5
Enter the measured value, and press Next.
CAL:CURR:LOW <reading>, (@<channel>)
6
Read DONE or FAIL on the display.
(wait 30 seconds)
SYST:ERR?
7
Press Cal Save.
CAL:SAVE
Save the calibration data
To save calibration data, go to the ‘CAL Save’ menu to save the calibration data or enable the ‘Auto Save’ feature.
With Auto Save, calibration data will be saved when the user exits the calibration menu.
After completing the Voltage, Current and Low Range Current calibrations, save the calibration data before exiting
the calibration state, or simply exit the calibration state if Auto Save is on.
To save the CAL data: CAL:SAVE
To enable the CAL Auto Save: CAL:ASAVE ON
To exit CAL State: CAL:STAT 1 <code>
46
Keysight E36300 Series Service Guide