Remote Interface Reference
3
SCPI Status Registers
E3632A User’s Guide
111
The Standard Event register is cleared when:
•
you execute the
*CLS
(clear status) command.
•
you query the event register using the
*ESR?
(Event
Status register) command.
For example,
28
(4 + 8 + 16) is returned when you have
queried the status of the Standard Event register, QYE,
DDE, and EXE conditions have occurred.
The Standard Event Enable register is cleared when:
•
you execute the
*ESE 0
command.
•
you turn on the power and have previously configured the
power supply using the
*PSC 1
command.
The enable register will not be cleared at power- on if you
have previously configured the power supply using the
*PSC
0
command.
4
EXE
16
Execution Error. An execution error occurred (refer to error messages
-211 to -223 in
“Execution Error Messages” on page 130
).
5
CME
32
Command Error. A command syntax error occurred (refer to error
messages -104 to -178 in
“Execution Error Messages” on page 130
).
6
Not Used
0
Always set to 0.
7
PON
128
Power On. Power has been turned off and on since the last time the
event register was read or cleared.
Table 3-5
Bit definitions — Standard Event register (continued)
Bit
Decimal value
Definition
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...