1
Calibration Procedures
40
Keysight E3632A Service Guide
To unsecure the power supply for calibration
The power supply can use a calibration security code to prevent unauthorized or
accidental calibration. This procedure shows you how to unsecure the power
supply for calibration from the front panel.
1
Turn on the front-panel calibration mode.
SECURED
Turn on the calibration mode by pressing the Calibrate key while
simultaneously turning on the power supply then continue to hold the
Calibrate key for about 5 seconds until a beep is heard.
If the power supply is secured, you will see the above message from the front
panel for approximately one second. The CAL MODE message is then
displayed on the front panel.
2
Move to the security code by pressing the Secure key.
000000 CODE
3
Enter the security code using the knob and resolution selection keys.
003632 CODE
The security code is set to “HP003632” when the power supply is shipped from
the factory. The security code is stored in non-volatile memory and does not
change when the power has been off or after a remote interface reset.
To enter the security code from the front panel, enter only the last six digits. To
enter the security code from the remote interface, you may enter up to 12
characters.
Use the resolution selection keys to move left or right between digits. Use the
knob to change the digits. Notice that the security code may be different if the
security code has been changed from the default setting.
4
Unsecure the power supply.
UNSECURED
The power supply is unsecured when you press the
Secure
key. You will see
the above message from the front panel for one second. The CAL MODE
message is displayed on the front panel after above message.
5
Turn off the calibration mode.
Turn off the power supply to exit the calibration mode.
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply Service Guide ...
Page 12: ...Keysight E3631A Service Guide 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...16 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 68: ...2 Service 70 Keysight E3632A Service Guide ...
Page 70: ...72 Keysight E3632A Service Guide Component locator diagram for the front panel top ...