Troubleshooting
1
Impedance Analyzer
33
Performance Test failure Troubleshooting
This section describes the adjustment and troubleshooting
procedures used when the Keysight E4991B fails the performance
tests. If the performance of the instrument is critical for the test
limits and seems adjustable, perform first the adjustment(s)
related to the failed test. When the test result are far from the
tolerance of the test or the performance is not adjustable, isolate
the faulty assembly in accordance with the
procedure.
Recommended Adjustment for Performance Test failure
shows the recommended adjustments when the
performance test fails. Select the adjustment program
corresponding to the recommended adjustment and perform the
adjustment.
Performance failure Troubleshooting
shows the recommended replacement board when the
performance test fails even after adjustment.
Table 2
Recommended adjustment for performance test failure
Failed Performance test item
Recommended Adjusment
Frequency Accuracy
Frequency Reference
Level Accuracy
Synthesizer Gain
Source Output Power
DC Bias Accuracy
DC Bias Level
Z Measurement Accuracy
Receiver IF Range
Default Cal Coefficient
Table 3
Recommended replacement board for performance test failure
Failed Performance test item
A21 Sour
ce
A6 R
eceiv
er
A9 DC Bias
A55 10MHz OCX
O
Test Head
Frequency Accuracy
**
*
Summary of Contents for E4991B
Page 1: ...Keysight E4991B Impedance Analyzer Troubleshooting Guide ...
Page 2: ......
Page 12: ...10 Network Analyzer ...
Page 17: ...Troubleshooting 1 Impedance Analyzer 15 Figure 2 Primary trouble isolation flowchart ...
Page 19: ...Troubleshooting 1 Impedance Analyzer 17 Figure 3 No display trouble isolation procedure ...