Replacing the Power-Line Fuse
Refer to
EDU36311A User's Guide
for details.
User Replaceable Parts
You can find the instrument support part list at Keysight's Test & Measurement Parts Catalog
http://www.keysight.com/find/parts
.
Battery Replacement
SHOCK HAZARD
Only qualified, service-trained personnel who are aware of the hazards involved should remove instrument
covers. Always disconnect the power cable and any external circuits before removing the instrument
cover. Some circuits are active and have power for a short time even when the power switch is turned off.
The internal battery powers the real-time clock. The primary function of the clock it to provide a time stamp for the
internal file system. If the battery fails, the clock and time stamp function will not be available. No other instrument
functions are affected.
Under normal use at room temperature, the lithium battery has a life expectancy between seven and ten years. Note
that battery life will be reduced if the instrument is stored for a prolonged period at temperatures above 40 degrees
Celsius.
The battery type is Panasonic CR 2032.
Replacing the Battery
1.
Remove the instrument front panel. See
.
2.
The battery is located at the main board.
3.
Press out the small spring clip. Use a flat-bladed plastic screwdriver and carefully pry up on the side of the bat-
tery.
10
Keysight EDU36311A Service Guide