Disassembly
Tools required
Items
Torque value
T25 Torx driver
21 in.lbs
T20 Torx driver
9.0 in.lbs
T10 Torx driver
7 in.lbs
Removing the front panel
Steps
Instructions
Visual
1.
On the top panel, remove two screws (M5x10) with T25 Torx driver from
the strap handle.
Then, remove the plastic cap and strap handle from the instrument.
2.
On the bottom panel, remove the rubber foot.
Remove four screws (M4x12) with T20 Torx driver from the rubber foot.
Then, remove the rubber foot from the instrument.
3.
Remove two screws (M3x6) with T10 Torx driver that holds the front
panel and the chassis cover.
4.
On the rear panel, remove three screws (M3x6) with T10 Torx driver.
5.
Pull out the chassis cover.
12