Performance Verification
Performance verification ensures that the instrument performs within the specifications stated in the data sheet
(
https://www.keysight.com/us/en/assets/3120-1003/data-sheets/EDU36311A-Triple-Output-Bench-Power-
)
Recommended test equipment
The test equipments recommended for the performance verification and adjustment procedures are listed below. If
the exact instrument is not available, use the accuracy requirements shown to select substitute calibration
standards.
Type
Specification
Recommended model
Digital multimeter
Readout: 8 1/2 digits Basic DC Accuracy: 0.0008%
Keysight 3458A or equivalent
Current monitor
15 A (0.1 Ω), TC = 4 ppm/°C
Guildline 9230A-15R
Electronic load
150 V, 5 A minimum, with transient capability and a slew rate of
833 kA/s or better.
Keysight N3300A mainframe, with
N330xA modules
LAN / USB controller
N/A
PC with Keysight Connection Expert
loaded
Oscilloscope
Sensitivity: 1 mV; Bandwidth Limit: 20 MHz Probe: 1:1 with RF
tip
Keysight DSOX1202A oscilloscope or
equivalent
Terminations
1 – 50 Ω BNC termination
2 – 50 Ω, ≥ 1/8 W resistor
N/A
Variable voltage transformer or
AC source
Adjustable to highest rated input voltage range. Power: 500 VA
Keysight 6813C or equivalent
Test considerations
– Ensure that the calibration ambient temperature is stable and between 20 °C and 30 °C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Keep cables as short as possible, consistent with the impedance requirements.
– Performance verification and calibration procedure must be performed through front panel output.
The tests should be performed by qualified personnel. During performance verification tests, hazardous voltages
may be present at the outputs of the power supply.
Keysight EDU36311A Service Guide