Calibration Adjustment Procedures
This chapter includes calibration adjustment procedures for Keysight EDU36311A power supply. Instructions are
applicable for performing the procedures from either the front panel or a controller over the LAN or USB.
Perform the verification tests before calibrating your instrument. If the instrument passes the verification
tests, the unit is operating within its calibration limits and does not need to be re-calibrated.
Closed–case electronic calibration
The instrument uses closed-case electronic calibration; no internal mechanical adjustments are required. The
instrument calculates correction factors based on reference signals that you apply and stores the correction factors
in non-volatile memory. This data is not changed by cycling power, or *RST.
Calibration interval
The recommended calibration interval for Keysight EDU36311A power supply is one year.
Calibration adjustment process
The following general procedure is recommended to complete a full calibration adjustment.
1.
Adhere to the test considerations. See
Performance Verification > Test considerations
for details.
2.
Perform the performance verification tests to characterize the instrument. See
for
details.
3.
Unsecure the instrument for calibration. See
for details.
4.
Perform the calibration procedures. See
for details.
5.
Secure the instrument against the calibration. See
for details.
6.
Take note of the security code and calibratizon count in the instrument's maintenance records.
7.
Perform the performance verification tests to verify the calibration.
Calibration security
The instrument has a calibration passcode to prevent accidental or unauthorized calibration. When you receive your
power supply, it is secured by a default passcode.
The default passcode is EDU36311A. The security code cannot be changed by a power cycle or *RST.
You can enter a passcode of up to 12 characters. The first character must be a letter (A-Z), remaining may contains
letters, numbers (0-9), or underscore "_". Blank spaces are not allowed.
You can change the passcode from both front panel and remote interface after unsecure the instrument using the
old passcode.
Keysight EDU36311A Service Guide
29