27.
Program the load settings as described in the test record form under “Resistance Programming - Low range,
Maximum resistance”. Program the power supply to the power supply settings described in the test record
form.
28.
Divide the voltage drop (DMM reading) across the current shunt by the shunt resistance to convert to amps and
record this value (Iin).
29.
Calculate the resistance value as follows: R = Vin/Iin. Vin is the voltage measured by the second DMM. Convert
the input resistance (R) to Siemens by inverting the value: S = 1/R. The value in Siemens should not exceed value
listed in the test record under “Resistance Programming - Low range, Maximum resistance”.
Keysight EL30000 Series Service Guide
21