Test Record Forms
Test record form - Keysight EL33133A
Test record form - Keysight EL34143A
Test record form - Keysight EL34243A
Test record form - Keysight EL33133A
Keysight EL33133A
Record Number: ______
Date: ______
Description
Lower Limits
Results
Upper Limits
Constant Voltage (CV) Tests
Voltage Programming & Readback - High range
Minimum voltage:
Voltage measured over interface:
Maximum voltage:
Voltage measured over interface:
Voltage Programming & Readback - Low range
Minimum voltage:
Voltage measured over interface:
Maximum voltage:
Voltage measured over interface:
1.48455 V
V
in
- 15.45 mV
149.94 V
V
in
- 60 mV
0.145755 V
V
in
- 4.245 mV
14.9913 V
V
in
- 8.7 mV
________
________
________
________
________
________
________
________
1.51545 V
V
in
+ 15.45 mV
150.06 V
V
in
+ 60 mV
0.154245 V
V
in
+ 4.245 mV
15.0087 V
V
in
+ 8.7 mV
Constant Current (CC) Tests
Current Programming & Readback - High range
Minimum current:
Current measured over interface:
Maximum current:
Current measured over interface:
Current Programming & Readback - Low range
Minimum current:
Current measured over interface:
Maximum current:
Current measured over interface:
0.49255 A
I
in
- 7.45 mA
39.9728 A
I
in
- 27.2 mA
0.049155 A
I
in
- 0.845 mA
3.99718 A
I
in
- 2.82 mA
________
________
________
________
________
________
________
________
0.50745 A
I
in
+ 7.45 mA
40.0272 A
I
in
+ 27.2 mA
0.050845 A
I
in
+ 0.845 mA
4.00282 A
I
in
+ 2.82 mA
Constant Resistance (CR) Tests
Resistance Programming - High range
Minimum resistance:
Maximum resistance:
Resistance Programming - Medium range
Minimum resistance:
Maximum resistance:
Resistance Programming - Low range
Minimum resistance:
Maximum resistance:
8.19 mS
0 S
83.9 mS
0 S
19.82 S
0 S
________
________
________
________
________
________
11.81 mS
2.05025 mS
116.1 mS
16.8008 mS
20.18 S
0.193367 S
To be continued on next page...
24