Keysight M809256PB OIF CEI-56G Rx Test Automation
Application
User Guide
8
OIF CEI-56G LR PAM4
Calibrations
Calibration Parameters in Debug Mode
LR configuration in the Test App
Receiver Interference Tolerance
This section provides the test procedures for the OIF CEI-56G Rx
Calibrations, which are applicable for Long Reach (LR) standards.
Before performing the OIF CEI-56G LR tests, you must calibrate all the
related parameters. Perform calibrations in the order displayed in the OIF
CEI-56G Rx Test Application.
As mentioned earlier, ensure that the FlexDCA Oscilloscope has been
calibrated and the mated Compliance Board is properly connected to the
testing instruments to perform OIF CEI-56G Calibrations.