Keysight M809256PB OIF CEI-56G Rx Test Automation Application User Guide
145
OIF CEI-56G LR PAM4 Calibrations
8
ID
Dependencies
This test depends on the values measured in “
• Test Channel S-Parameter
• Noise Channel S-Parameter
• Excel Config File—containing COM parameter values after “Transmitter
Measurements for COM Model is performed”
For more information on configuring Test Channel S-Parameter, Noise
Channel S-Parameter and Excel Config File, refer to “
Parameters
Refer to “
Calibration Parameters in Debug Mode
Procedure
This calibration determines the Broadband Noise (RMS) value using COM
calibration until the COM value begins decreasing below 3dB.
Results
This calibrated values are displayed in a tabular format in the Results tab.
Figure 115
Rx Calibration using COM Model results in tabular format
Standard Name
Test ID
Long Reach
671106