160
Keysight M8194A 120 GSa/s Arbitrary Waveform Generator User Guide
5
M8194A Maintenance
Power and Ventilation Requirements
For power and ventilation requirements, refer to:
•
http://www.keysight.com/find/M9514A
for 14-slot chassis related
documentation.
•
http://www.keysight.com/find/M9505A
for 5-slot chassis related
documentation.
•
http://www.keysight.com/find/M9502A
for 2-slot chassis related
documentation.
Thermal Protection
Overheating Detection
The instrument monitors its internal temperature. If the temperature
exceeds approximately 80°C the power supply is switched off. The
instrument will not turn on automatically if the temperature is decreasing
again.
Fan Failure
If a fan is broken or prevented from operating by a blockage the
temperature will increase. When the temperature exceeds approximately
80°C the overheating detection switches off the instrument for safety
reasons. For reliability it is recommended to send instruments with broken
or defective fans immediately to Keysight Service for repair.
Battery
The M8194A module does not have a battery.
Operating Environment
For details on operating environment for M8194A module, refer to the
section