Keysight M9018A PXIe 18-Slot Chassis Startup Guide
STEP 3: Turn On and Verify Operation of the Chassis System
PC startup requirements for Keysight IO Libraries Suite
Restarting of the PC after Windows is up ensures that Connection Expert 16.0
will recognize the chassis. Note the need, before doing the restart, to wait a
sufficient length of time after Windows comes up if the chassis configuration has
changed. This wait time is also required on the first-ever boot of the PC with of
the chassis.
If Connection Expert 16.0 ever fails to recognize the chassis or appears to have
an incorrect view of the chassis modules, the PC should be restarted using the
sequence shown in Figure 6.
While Figure 6 depicts a system with a remote controller PC, the same
requirement applies to a chassis containing an embedded controller. After
powering up the chassis (which powers up the embedded controller), the
embedded controller should be restarted.
PC startup events for Connection Expert 16.1 or later
With IO Libraries Suite 16.1 or later, the need for the restart shown in Figure 6
has been reduced. However, there is still a need, before starting Connection
Expert, to wait a sufficient length of time after Windows comes up if the chassis
configuration has changed. This wait time is also required on the first-ever boot
of the PC with of the chassis. This sequence is shown in Figure 7.
Figure 7
Startup sequence for Connection Expert 16.1 or later
Boot
#1
Chassis power
(High = powered up)
If the chassis configuration has changed since the last
boot, sufficient time needs to be allowed between
when Windows is up and when t Connection
Expert 16.1 is started to ensure that Windows is able to
assign drivers to the new modules that it finds This is
Remote Controller PC
(High = turned on)
assign drivers to the new modules that it finds. This is
usually completed within one minute, but can take
longer depending on the number of new modules, the
speed of your computer, and other software being
executed as the computer boots up.
Remote Controller PC
Startup Event(s)
On subsequent boots and presuming that the
configuration hasnt changed since the first boot,
onnection Expert can be started immediately
after Windows is up because Windows has kept a copy
f h d i
i
d f ll
i
h fi
b
d
Windows is up as indicated
by the presence of the
Wi d
l
h
Starting of
Connection Expert 16.1
of the drivers assigned following the first boot, and uses
this information to quickly assign drivers following
subsequent boots.
Windows splash screen.
Even with IO Libraries 16.1 and later versions, it is always advisable to boot the
PC a second time to ensure that the PC properly enumerates the chassis.