Performance Verification Tests
Specification
Table 1
Insertion Loss Conditions
Source
≤ 2 dB
RF IN to RF 1 OUT, 100 kHz to 2.9 GHz, minimum attenuation
≤ 6 dB
RF IN to RF 1 OUT, 2.9 GHz to 26.5 GHz, minimum attenuation
≤ 7 dB
RF IN to RF 2 OUT (Bypass), minimum attenuation
≤ 1 dB
LO IN to LO 1 OUT, ≤ 10 GHz
≤ 1.5 dB
LO IN to LO 1 OUT, ≤ 26.5 GHz
≤ 1 dB
LO IN to LO 2 OUT, ≤ 10 GHz
≤ 1.5 dB
LO IN to LO 2 OUT, ≤ 26.5 GHz
Equipment
Table 2
Equipment
Supported Models
Notes
Vector Network
Analyzer (VNA)
Agilent N5245A
Other models include N5244A; N5230C Opt 525; N5230C Opt 520;
N5230C Opt 425; N5230C Opt 420; E8361C; E8362C; E8363C;
E8364C. If you use one of these models, download the uncertainty
calculator to calculate (worse-case) the uncertainty values for the
measurements:
www.agilent.com/find/na_calculator
.
Calibration Kit
Agilent N4691B ECal Opt
OOF; 85052B
Cables
85133D cable kit
The supported models listed here assume N5245A. The VNAs
specified are configured with 2.4 mm and 1.85 mm. Appropriate
cables and adaptors are needed to connect to the M9360A 3.5 mm
female connectors.
Cable
8121-1221
Coaxial cable, 3.5 mm (m) connectors, 1220 mm length
Adaptor
83059A
Coaxial adaptor, 3.5 mm (m) to 3.5 mm (m)
Adaptor
11901C
Adaptor, 2.4 mm (m) to 3.5 mm (f)
Adaptor
11901D
Adaptor, 2.4 mm (f) to 3.5 mm(m)
Test Configuration
24
Agilent M9360A PXI Attenuator/Preselector Service Guide
Summary of Contents for M9360A
Page 1: ...Service Guide Keysight M9360A PXI Attenuator Preselector 100 KHz to 26 5 GHz ...
Page 2: ......
Page 7: ......