M9383A Startup Guide
Safety and Maintenance Information
Safety Information
Safety Information
IMPORTANT
The safety of any system incorporating the equipment is the responsibility of
the assembler of the system.
IMPORTANT
Proper Ergonomics should be considered when using accessories such as a
keyboard or a mouse.
IMPORTANT
When installing the instrument(s) into a cabinet, consideration shall be given
to the convection flow into and out of the cabinet. Consideration shall also be
given to the individual instruments to avoid having the heated discharge of
one instrument, now becoming the cooling intake air for another instrument.
Another area of concern is verification that the maximum ambient operating
temperature of the instrument(s) is not exceeded by cabinet installation.
Keysight recommends forced air convection whenever an instrument(s) are
installed in a cabinet and further recommends that the maximum operating
temperature of the cabinet be reduced 10°C from the lowest, of the maximum
operating temperature of a single instrument. If there are any concerns or
special requirements a Keysight Field Engineer should be consulted to assure
instrument(s) temperature compliance and performance.