12
Keysight MIPI M-PHY Frame Generator User Guide
2
Test Instrument Setup
The advantage of the DSGA setup compared to M8040A alone
comes from the low-speed pattern analyzer. When the link training is
completed, the DUT is put into test mode and the output of the DUT is
analyzed by the DSGA to check that the link training has been successful.
Since PWM data is generated with the M8040A by bit multiplication - and
depending on the test conditions the M8040A can be running at high
speed Gears - the generation of the training sequence by the M8040A will
take much longer than with the DSGA.
The DSGA error detector requires the DUT responses to come in PWM
mode, either as bursts or continuous.
Also in this system configuration, either a Keysight M8045A-803 matched
coupler pair or a 3dB attenuator should be connected to the M8057
remote head output of the M8045 module. We recommend using a 3dB
attenuator with a minimum frequency range of 26.5GHz, 2 pieces for each
data lane.
BIT-3000 DSGA MIPI
®
M-PHY Generator
To use this setup, select “Bit3000” as the “Instrument” in the Connection
window of the software. For more details, see
This setup can be used to perform transmitter tests using the Keysight
scopeapp application.
The DSGA (BIT-3000), see
, can be used alone to generate the
signal required to put the DUT into test mode such that it starts
transmitting the required signal with which the Tx tests at different data
rates can be performed with the help of the Keysight scopeapp
application.
Summary of Contents for MIPI M-PHY
Page 1: ...User Guide Keysight MIPI M PHY Frame Generator ...
Page 5: ...Keysight MIPI M PHY Frame Generator User Guide 1 Introduction Overview 6 ...
Page 8: ...8 Keysight MIPI M PHY Frame Generator User Guide 1 Introduction ...
Page 18: ...18 Keysight MIPI M PHY Frame Generator User Guide 3 Software Installation and Update ...
Page 22: ...22 Keysight MIPI M PHY Frame Generator User Guide 4 Starting and Registering the Software ...
Page 70: ...70 Keysight MIPI M PHY Frame Generator User Guide 6 Software Installation and Update ...
Page 74: ...72 Keysight MIPI M PHY Frame Generator User Guide 7 Software Installation and Update ...
Page 75: ...Keysight MIPI M PHY Frame Generator User Guide 75 ...