InfiniiMax RC Probe Amplifier and Probe Heads User’s Guide
11
Configuring Infiniium Software for Probe Amplifier and Probe Heads
d
For the solder-in and ZIF probe heads, adjust the shape of the probe head
tips as shown in the following figure before you clamp these to the deskew
fixture. This is needed to ensure a proper contact with the deskew fixture.
e
Use the N2787A 3D Probe Positioner to hold the probe in place and to
maintain a steady contact of probe tips with the deskew fixture.
f
While pressing down on the MX0104A fixture’s spring-loaded clamps, insert
the probe head tips beneath the clamps. Place the positive tip of the probe
head on the center conductor of the deskew fixture. The negative tip or
ground lead of the probe head must be connected to the ground plane on
either side of the center conductor of the deskew fixture. The MX0104A’s
NOTE
Make sure that the probe head is approximately
perpendicular to the deskew fixture.