62
InfiniiMax RC Probe Amplifier and Probe Heads User’s Guide
6
MX0100A InfiniiMax Micro Probe Head
To check the MX0100A probe head for any damage
You can use a Digital Multimeter to check the resistance measurement of your
MX0100A probe head. If the resistance measurement between the probe head’s tip
and the center conductor of the SMP connector is ~25.2 kohm, then the probe
head is usable.
Figure 24
Resistance measurement for an undamaged MX0100A probe head
For a damaged probe head, the resistance measurement between the MX0100A
probe head tip wire and the center conductor of the SMP connector of the
MX0100A is displayed as Infinite.
Figure 25
Resistance measurement for a damaged MX0100A probe head