48
Keysight N437x Series Lightwave Component Analyzer, Programmer’s Guide
1
Remote Operation
Example
:NWA:MOD? -> N5245A
Command
:NWA:PORT:NUMBer?
Syntax
:NWA:PORT:NUMBer?
Description
The number of network analyzer ports
Parameters
none
Response
integer
C#
(property) NumNWAPorts
Example
:NWA:PORT:NUMBer? -> 4
Command
*OPC?
Syntax
*OPC?
Description
Retrieves the operation complete state
Parameters
none
Response
<string> 0 | 1
C#
(method) OperationComplete()
Example
*OPC? -> 1
Command
:PARameter:ADVAnced:MODE?
Syntax
:PARameter:ADVAnced:MODE?
Description
Returns 1 if advanced mode is enabled. In advanced mode you can force
the LCA to switch the laser on or off, independent of the measurement
type. You also have additional optical- and RF- path de-embedding
possibilities, or can apply additional de- embedding on the receiver and
the source side, independent of the measurement type.
Parameters
none
Response
C#
<string> 0 | 1 (property) Advanced
Example
:PAR:ADVA:MODE? -> 0