Installation Note N5247-90113 65
Installation Procedure for the Upgrade
Figure 33
EEPROM Backup Menu
Operator’s Check
Perform the Operator’s Check to check the basic functionality of the analyzer.
For instructions, click the Chapter 3 bookmark “Tests and Adjustments” in the
PDF Service Guide
If you experience difficulty with the basic functioning of the analyzer, contact
Keysight. Refer to
“Contacting Keysight” on page 5
Calibration
Although the analyzer functions, its performance relative to its specifications
has not been verified. It is recommended that a full instrument calibration be
performed using the analyzer’s internal performance test software. To view
information on the performance test software, click the Chapter 3 bookmark
“Tests and Adjustments” in the PDF Service Guide
.
Step 35. Prepare the PNA for the User
1. If necessary, reinstall front jumper cables.
2. Install the cable guards, pushing them over the front jumper cables until
the cushioning material touches the front panel of the PNA.
3. Install the dust caps on the test ports.
4. Clean the analyzer, as needed, using a damp cloth.
Summary of Contents for N5247-60113
Page 3: ...3 ...
Page 4: ...4 ...
Page 66: ...66 Installation Note N5247 90113 Installation Procedure for the Upgrade ...
Page 67: ...Installation Note N5247 90113 67 Installation Procedure for the Upgrade ...
Page 68: ...68 Installation Note N5247 90113 Installation Procedure for the Upgrade ...