Verification and Calibration
Appendix B
Series N5700 User’s Guide
113
Test Record Form – Keysight N5751A and N5771A
Keysight N5751A and N5771A
Report No _______________
Date __________________
Description
Model
Minimum Specs.
Results
Maximum Specs.
Voltage Programming & Readback
Minimum Voltage Vout
Measurement Readback
High Voltage Vout
Measurement Readback
Both
Both
Both
Both
600 mV
Vout
−
300 mV
299.7 V
Vout
−
600 mV
__________
__________
__________
__________
900 mV
Vout + 300 mV
300.3 V
Vout + 600 mV
CV Load Effect
Both
−
32 mV
__________
+ 32 mV
CV Source Effect
Both
−
32 mV
__________
+ 32 mV
CV Ripple and Noise
peak-to-peak
rms
Both
Both
N/A
N/A
__________
__________
150 mV
20 mV
Transient Response
Voltage @ 2ms
Both
−
1 V
__________
+ 1 V
Current Programming & Readback
Minimum Current Iout
Measurement Readback
High Current Iout
Measurement Readback
N5751A
N5771A
N5751A
N5771A
N5751A
N5771A
N5751A
N5771A
10 mA
20 mA
Iout
−
7.5 mA
Iout
−
15 mA
2.495 A
4.99 A
Iout
−
10 mA
Iout
−
20 mA
__________
__________
__________
__________
__________
__________
__________
__________
15 mA
30 mA
Iout + 7.5 mA
Iout + 15 mA
2.505 A
5.01 A
Iout + 10 mA
Iout + 20 mA
CC Load Effect
N5751A
N5771A
−
5.5 mA
−
6 mA
__________
__________
+ 5.5 mA
+ 6 mA
CC Source Effect
N5751A
N5771A
−
2.25 mA
−
2.5 mA
__________
__________
+ 2.25 mA
+ 2.5 mA
WARNING
Return the voltage and current settings to zero when verification is completed
Test Description
N5751A Setting
N5771A Setting
Voltage Programming & Readback, Min Voltage
750 mV, 12.5 mA
750 mV, 25 mA
Voltage Programming & Readback, High Voltage
300 V, 12.5 mA
300 V, 25 mA
CV Load Effect, Source Effect, Ripple and Noise
300 V, 2.5 A
300 V, 5 A
Transient Response
200 V, from 0.25 A to 2.25 A
200 V, from 0.5 A to 4.5 A
Current Programming & Readback, Min Current
12.5 mA, 300 V
25 mA, 300 V
Current Programming & Readback, High Current
2.5 A, 300 V
5 A, 300 V
CC Load Effect, Source Effect
2.5 A, 300 V
5 A, 300 V
Load Requirements
N5751A
N5771A
Current shunt
0.1
Ω
15 A
0.1
Ω
15 A
Use fixed resistor instead of load modules
120
Ω
1kW
60
Ω
2kW
Fixed Resistor for CV Ripple and Noise
120
Ω
1kW
60
Ω
2kW