32
N7010A User’s Guide
3
Performance Verification
14
On the oscilloscope, press
[Clear Display]
and locate the oscilloscope’s Mean V
rms
reading, which is labeled “Mean” in the Measurements display area at the bottom
of the screen. Record this value here and in
out
at 50 MHz.
V
out
at 50 MHz: ______________
15
Calculate the reference gain at 50 MHz using the following expression. Record this
value here and in
on page 37 as the Calculated Gain at 50 MHz.
Gain at 50 MHz: ______________
16
Change the microwave generator’s frequency to 30 GHz.
17
Change the oscilloscope’s horizontal scale to 50 ps/div as shown in
Figure 14
Change Horizontal Scale to 50 ps/division
18
The power meter reads the input power to the scope channel. Convert this
measurement to V
rms
using the following expression. For example, if the power
meter reading is 4.0
μ
W, then V
in
= (4.0 x 10
-6
x 50
Ω
)
1/2
= 14.1 mV
rms
.
Gain
50 MHz
V
out @50 MHz
V
in @50 MHz
-----------------------------------
=
V
in
P
meas
50
Ω
×
=