Verification and Calibration
Appendix B
Series N8700 User’s Guide
117
Test Record – Keysight N8741A [300V, 11A, 3.3kW]
Keysight N8741A
Description
Report No _______________
Date __________________
Minimum Specs.
Results
Maximum Specs.
Voltage Programming & Readback
Minimum Voltage Vout
Measurement Readback
High Voltage Vout
Measurement Readback
600 mV
Vout
−
300 mV
299.7 V
Vout
−
600 mV
_____________
_____________
_____________
_____________
900 mV
Vout + 300 mV
300.3 V
Vout + 600 mV
CV Load Effect
−
50 mV
_____________
+ 50 mV
CV Source Effect
−
32 mV
_____________
+ 32 mV
CV Ripple and Noise
peak-to-peak
rms
N/A
N/A
_____________
_____________
300 mV
100 mV
Transient Response
Voltage @ 2 ms
−
1.5V
_____________
+ 1.5V
Current Programming & Readback
Minimum Current Iout
Measurement Readback
High Current Iout
Measurement Readback
44 mA
Iout
−
33 mA
10.967 A
Iout
−
44 mA
_____________
_____________
_____________
_____________
88 mA
Iout + 33 mA
11.033 A
Iout + 44 mA
CC Load Effect
−
7.2 mA
_____________
+ 7.2 mA
CC Source Effect
−
3.1 mA
_____________
+ 3.1 mA
WARNING
Return the voltage and current settings to zero when verification is completed
Test Description
N8741A Settings
Voltage Programming & Readback, Min Voltage
750 mV, 66 mA
Voltage Programming & Readback, High Voltage
300V, 66 mA
CV Load Effect, Source Effect, Ripple and Noise
300V, 11A
Transient Response
200V, from 1.1A to 9.9A
Current Programming & Readback, Min Current
66 mA, 300V
Current Programming & Readback, High Current
11A, 300V
CC Load Effect, Source Effect
11A, 300V
N8741A Load Requirements
Current shunt
0.1
W
15 A
Use fixed resistor instead of load modules
27.3
W
3.5 kW
(or Amrel 5KW 600V 300A electronic load)
Fixed Resistor for CV Ripple and Noise
27.3
W
3.5 kW