5
Language Reference
76
Series N8700 User’s Guide
Note that triggered values can be programmed outside these limits,
but an error will be generated when the trigger occurs.
Model (V rating)
8V
10V
15V
20V
30V
40V
60V
80V
100V
150V
300V
600V
Minimum voltage
0
0
0
0
0
0
0
0
0
0
0
0
Maximum voltage
8.4
10.5
15.75
21
31.5
42
63
84
105
157.5
315
630
[SOURce:]VOLTage:LIMit:LOW <value>|MIN|MAX
[SOURce:]VOLTage:LIMit:LOW? [MIN|MAX]
This command sets the low voltage limit of the output. When a low
voltage limit has been set, the instrument will ignore any
programming commands that attempt to set the output voltage below
the low voltage limit. The*RST value = Min.
The range of values that can be programmed for this command is
coupled with the immediate voltage level setting. The maximum value
for the low voltage limit is either the value in the following table, or
the immediate voltage setting multiplied by 0.95; whichever is lower.
The minimum setting is the value in the table.
Model (V rating)
8V
10V
15V
20V
30V
40V
60V
80V
100V
150V
300V
600V
Minimum low limit
0
0
0
0
0
0
0
0
0
0
0
0
Maximum low limit
7.6
9.5
14.25
19
28.5
38
57
76
95
142
285
570
[SOURce:]VOLTage:PROTection:LEVel <value>|MIN|MAX
[SOURce:]VOLTage:PROTection:LEVel? [MIN|MAX]
This command sets the over-voltage protection (OVP) level of the
output. The values are programmed in volts. If the output voltage
exceeds the OVP level, the output is disabled and OV is set in the
Questionable Condition status register. The*RST value = Max.
The range of values that can be programmed for this command is
coupled with the immediate voltage level setting. The minimum value
for the voltage protection level is either the value in the following
table, or the immediate voltage setting multiplied by 1.05; whichever
is higher. The maximum setting is the value in the table.
An over-voltage condition can be cleared with the Output Protection
Clear command after the condition that caused the OVP trip is
removed.
Model (V rating)
8V
10V
15V
20V
30V
40V
60V
80V
100V
150V
300V
600V
Min. protection limit
0.5
0.5
1.0
1.0
2.0
2.0
5.0
5.0
5.0
5.0
5.0
5.0
Max. protection limit
10
12
18
24
36
44
66
88
110
165
330
660