28
Installation Note N9020-90257
Option BBA, Analog Baseband I/Q Inputs Retrofit Kit
7.
The MXA will automatically consume the License File (this may take a few minutes). When the
License File is consumed the Keysight License Manager will display a “
Successful
License Installation
” message similar to the one shown in
.
Figure 20
Successful License Installation
Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following folder on the MXA C:\Program Files\Agilent\licensing.
Verify the Installation
1.
If you haven’t already verified the BBA hardware and license, press
System
,
Show System
on
the analyzer to display a list of all installed options. You should notice the BBIQ boards and
license.
Before performing adjustments and performance verification tests using the N7800A Calibration
and Adjustment Software for Keysight RF and Microwave Instruments, press
Input/Output
,
I/Q
,
I/Q
,
I/Q Cable Calibrate...
on the analyzer and follow the on-screen directions.
If the newly installed hardware is functioning properly the cable cal should work fine.