4
Installation Note
N9030-90045
Option BBA, Analog Baseband I/Q Inputs Retrofit Kit
Introduction
This kit provides all parts and instructions necessary to install Option BBA, Analog Baseband I/Q
Input, into the Agilent N9030A with frequency range 26.5 GHz and below.
NOTE
1. The signal analyzer requires software revision A.08.54 or newer.
This retrofit kit installation requires that you re-install instrument software even if the current
software meets the revision requirement above.
The latest revision of the PXA Signal Analyzer software may be downloaded from
http:/www.agilent.com/find/xseries_software.
The instrument software must be installed AFTER the hardware in the kit is installed. Follow the
procedure in this document for redeeming and installing the option upgrade license key.
2. The option is licensed for one instrument model/serial number combination. The license key will
only install on the designated instrument.