Specifications
Table 7. 85309B specifications
.........................................................................................................................
85309B Options 40x
Output Power Limits
for HIGH Band
LO Output Power (+6 dBm Input)
Frequency Range
Power Level
0.3 – 0.5 GHz
> +21.3 dBm
0.5 – 3 GHz
> +22.4 dBm
3 – 6.2 GHz
> +24.4 dBm
6.2 – 18 GHz
> +22 dBm
LO Output Power (0 dBm Input)
0.3 – 0.5 GHz
> +21.3 dBm
0.5 – 3 GHz
> +22.4 dBm
3 – 6.2 GHz
> +24.4 dBm
6.2 – 18 GHz
> +22 dBm
85309B Standard or
Options 001 or 002
Output Power Limits
LO Output Power (+6 dBm Input)
Frequency Range
Power Level
0.3 – 0.5 GHz
> +21.6 dBm
0.5 – 3 GHz
> +24.1 dBm
3 – 6.2 GHz
> +26.3 dBm
6.2 – 18 GHz
> +23.5 dBm
LO Output Power (0 dBm Input)
0.3 – 0.5 GHz
> +21.6 dBm
0.5 – 3 GHz
> +23 dBm
3 – 6.2 GHz
> +25 dBm
6.2 – 18 GHz
> +23 dBm
85309B Options 40x
Output Power Limits
for LOW Band
LO Output Power (+6 dBm Input)
Frequency Range
Power Level
0.1 – 1 GHz
> +14 dBm
LO Output Power (0 dBm Input)
0.1 – 1 GHz
> +13 dBm
Remote mixer distances
Mixers require a certain LO drive power level; the output power of the 85309B LO/IF dis-
tribution unit and the RF loss of the cables will determine the maximum allowable cable
lengths. Maximum cable lengths can be calculated using the following equations:
Cable (source to 85309B) length (meters) = (P
OUT
source – P
IN
85309B)/(cable loss/
meter at frequency)
Cable (85309B to mixers) length (meters) = (P
OUT
85309B – P
IN
mixer)/(cable loss/meter
at frequency)
55 | Keysight | Antenna Test – Selection Guide