4
Service and Maintenance
52
Keysight U1241AN User’s and Service Guide
Fuse Replacement
1
Turn the multimeter off and disconnect the test leads from external
equipment.
2
Loosen the four screws at the rear panel, lift and remove the cover.
3
Gently remove the defective Fuse 1 by prying one end of the fuse and removing
it out of the fuse bracket, see
4
Replace a new fuse of the same size and rating into the center of the fuse
holder.
Figure 4-2
Fuse 1 replacement
5
If you want to replace a defective Fuse 2, remove Fuse 1 prior to removing the
Fuse 2. Loosen the four screws (shown in
on page 53) to lift and
remove the circuit board from top case.
6
Gently remove the defective Fuse 2 by prying one end of the fuse loose and
removing it out of the fuse bracket, see
7
Replace a new fuse of the same size and rating into the center of the fuse
holder.
NOTE
You are recommended to use clean/dry gloves when performing fuse
replacement. Do not touch any components except the fuse and plastic parts.
No recalibration is required after replacing the fuse.
Fuse 1 (440 mA/1000 V)