5
Performance Tests and Calibration
70
Keysight U1241AN User’s and Service Guide
Unsecuring the multimeter without the security code
1
Record the last 4 digit serial numbers of the multimeter.
2
Turn the rotary switch to
.
3
Press
and
simultaneously to enter the Calibration Security Code
entry mode. The primary display shows
5555
and the secondary display shows
SEC
.
4
Press
for more than one second to enter Set Default Security Code
mode. The secondary display shows
SEr
and primary display shows “
5555
”.
5
Press
or
to step each character in the code. Press
and
to
change the value of the selected character.
6
Set the last 4 digit serial number of the multimeter. Press
(Save) to
confirm the entry.
7
If the 4 digit serial number entered is correct, the secondary display will show
PAS
. If an invalid code is entered, the multimeter will shows error code
E03
.
Ensure that the last 4 digit serial number is entered correctly and repeat the
step 1 to 7.
Using the Front Panel for Adjustments
This section describes the process used to perform adjustments from the front
panel.
Selecting the Adjustment Mode
Unsecure the multimeter, see
“Unsecuring the Multimeter for Calibration”
page 68 or
“Unsecuring the multimeter without the security code”
Once the multimeter has been unsecured, the reference value will be indicated on
the primary display.
Summary of Contents for U1241AN
Page 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Page 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...