5
Performance Tests and Calibration
72
Keysight U1241AN User’s and Service Guide
Valid Adjustment Input Values
Adjustment can be accomplished using the following input values below.
CAUTION
Do not power off the multimeter during adjustments, as this may delete the
calibration memory for the present function.
Table 5-3
Valid adjustment input values
Function
Range
Valid Input Reference Values
1000 mV, 10 V, 100 V, 1000 V
0.9 to 1.1 x full scale
1000 mV, 10 V, 100 V, 1000 V
0.9 to 1.1 x full scale
1 V
0.9 to 1.1 x full scale
1000
Ω
, 10 k
Ω
, 100 k
Ω
,
1000 k
Ω
, 10 M
Ω
0.9 to 1.1 x full scale
1000 nF, 10
μ
F, 100
μ
F, 1000
μ
F, 10 mF
0.9 to 1.1 x full scale
1000
μ
A, 10000
μ
A
0.9 to 1.1 x full scale
100 mA, 1000 mA
0.9 to 1.1 x full scale
10 A
0.9 to 1.1 x full scale