5
Performance Tests and Calibration
76
Keysight U1241AN User’s and Service Guide
4
Turn the rotary switch to
position
Short
Dual banana plug with
copper wires short
between two terminals
SHrt
10 M
Ω
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
oPEn
10 M
Ω
10.000 M
Ω
1000 k
Ω
1000 k
Ω
1000.0 k
Ω
100 k
Ω
100 k
Ω
100.00 k
Ω
10 k
Ω
10 k
Ω
10.000 k
Ω
1000
Ω
1000
Ω
1000
Ω
5
Turn the rotary switch to
position
Open
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
oPEn
1000 nF
400 nF
400.0 nF
1000 nF
1000.0 nF
10
μ
F
10
μ
F
10.000
μ
F
100
μ
F
100
μ
F
100.00
μ
F
1000
μ
F
1000
μ
F
1000.0
μ
F
10 mF
10 mF
10.000 mF
6
Turn the rotary switch to
position
Open
Input terminals open
(remove all test leads and
shorting plugs from input
terminals)
oPEn
1000
μ
A
1000
μ
A
1000.0
μ
A
10000
μ
A
10000
μ
A
10000
μ
A
Table 5-4
Calibration adjustments (continued)
Step
Test function
Cal range
Input reference value
Cal item
U1241AN
Summary of Contents for U1241AN
Page 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Page 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...