Performance Tests and Calibration
62
Keysight U1241AN User’s and Service Guide
Calibration Process
1
Prior to performing the verification tests, see
2
Perform the verification tests to characterize the multimeter, see
3
Unsecure the multimeter for calibration, see
4
Prior to performing the adjustments, see
5
Perform the adjustment procedures, see
6
Secure the multimeter against unauthorized calibration, see
on page 78. Ensure that the multimeter has quit the
Adjustment Mode and turned off.
7
Record the new security code and calibration count in the multimeter’s
maintenance records.