Performance Test Calibration
2
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
19
DC current gain verification test
DC current configuration:
(
CONFigure:CURRent[:DC] [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
1
Connect the calibrator to the front panel current HI and
LO input connectors.
2
Select each function and range based on the input current
in
Table 2- 4
.
3
Make a measurement and observe the result. Compare the
measurement results with the appropriate test limits
shown in
Table 2- 4
. (Be certain to allow for appropriate
source settling time when using the Fluke 5520A.)
[1]
Select NPLC 20 digit resolution.
Q
= Quick performance verification test points
C A U T I O N
Set the calibrator output to 0 V before disconnecting it from the
multimeter input terminals
Table 2-4
DC current gain verification
Input
Function
[1]
Range
Quick check
Error from nominal 1
year
10 mA
DC Current
10 mA
Q
± 7.5 µA
100 mA
100 mA
± 65 µA
1 A
1 A
±1.57 mA
2 A
2 A
Q
±3.14 mA