U3042AE16
Controlling the Test Set with N5222/32/42A/B
RF Path Configuration with Option 029
If the analyzer has Option 029 or H29 (noise figure capability) verify that the path configuration
is set to Default. If it is not in Default configuration, Port 5 will not operate correctly.
1. Select
Trace/Chan > Channel > Hardware Setup > Path Cong...
-OR-
(Instrument >
Setup >
Internal Hardware > RF Path Config...)
and in the drop-down
menu select
Default
>
OK
. Save this configuration as a User Preset by selecting
Save
>
User Preset > Save current state as User Preset
.
Do not use the factory Preset (User Preset Off), the analyzer will return to Option 029
path configuration.
2. Verify that the Port 1 Noise Tuner Switch is set to external. Select
Utilities System
>
Configure
>
Preferences
-OR-
(Utility > System > System Setup > Preferences > User
Preset)
.
If not, select
Meas: Port 1 Noise Tuner Switch is set to external
.
Figure 3
5
RF Path Configuration
User's and Service Guide U3042-90012
37