U3042AE16
Cal Kit Operational Check
Cal Set Verification
1. Select
Trace/Chan
>
Trace
>
Delete Trace
-OR-
(Instrument > Trace > Delete Trace)
.
There should be no traces on the screen.
2. To launch the Cal Set Viewer toolbar. Select
Response > Cal > Manage Cals > Cal Set Viewer
-OR-
(Response > Cal > Cal Sets & Cal Kits > Cal Set Viewer [On/Off])
.
Figure
51
Calibration Cal Set Viewer
3.
From the Cal Sets drop-down menu, select
999.1
and select
Enable
. Select the
Reflection
Tracking(x,x)
, where x,x is the port being tested. Ensure that the
Enable
and
Error
Terms
check boxes are selected.
4.
Compare the Reflection Tracking (1,1) trace to the appropriate limits in
or
. Use the Markers function to measure the trace values.
5.
Repeat
for Cal Sets “999.1” thru “999.20” (20-Port).
60
User's and Service Guide U3042-90012